Benjamin HarrisSep 14, 2025
What to do if KaVo Dental equipment display shows ID 75?
- TTonya RobinsonSep 14, 2025
If the KaVo dental equipment displays 'ID 75', it means that Centramat is too full. Notify a service technician.
What to do if KaVo Dental equipment display shows ID 75?
If the KaVo dental equipment displays 'ID 75', it means that Centramat is too full. Notify a service technician.
Why there is no heating function on the multifunctional handpiece of KaVo Dental equipment?
If there's no heating function on the multifunctional handpiece of your KaVo dental equipment, ensure that spray heating is preselected.
Why the LED on the 'LP/AP' button flickers on KaVo Dental equipment?
If the LED on the 'LP/AP' button is flickering on the assistant element of your KaVo dental equipment, the data connection to the foot control is faulty. First, turn the wireless foot control on. If the problem persists, call a service technician.
Why is there no cold light on my KaVo ESTETICA E50 Life Dental equipment multifunctional handpiece?
If there is no cold light on your KaVo dental equipment multifunctional handpiece, the heating function may be preselected. Preselect cold light instead.
Why is the LED on the "LP/AP" button flickering on my KaVo Dental equipment assistant unit?
If the LED on the "LP/AP" button is flickering on your KaVo dental equipment assistant unit, the data connection to the foot control is faulty. Turn the wireless foot control on. If the problem continues, call a service technician.
Why is the LED on the "AP1" button flickering on my KaVo ESTETICA E50 Life Dental equipment dentist unit?
If the LED on the "AP1" button is flickering on your KaVo dental equipment dentist unit, it indicates that the data connection to the assistant unit is faulty. Turn the wireless foot control on. If the problem continues, call a service technician.
Why does my KaVo Dental equipment make a signal sound every 10 seconds and the LED flashes?
If your KaVo dental equipment is issuing a signal sound every 10 seconds and the LED on the 'Intensive germ reduction' key flashes green, the Oxygenal container is empty. Refill the Oxygenal container.
How to replace a defective high-pressure lamp on KaVo Dental equipment?
Replace the high-pressure lamp or Multi LED on the handpiece.
What to do if KaVo Dental equipment display shows ID 74?
If the KaVo dental equipment displays 'ID 74', and the indicator shows that the DEKASEPTOL bottle is not full after 600 seconds (10 minutes), it means that Centramat is empty. Notify a service technician.
Why there is no cold light on the handpieces of KaVo Dental equipment?
If there is no cold light on the handpieces of your KaVo dental equipment, first ensure that cold light is preselected. If that's not the issue, the high-pressure lamp or LED lamp on the handpiece may be defective and require replacement. Refer to the Instructions for Use of the handpiece for guidance.
Type | Dental treatment unit |
---|---|
Model | ESTETICA E50 Life |
Manufacturer | KaVo |
Integrated endodontics functions | Yes |
Integrated surgical functions | Yes |
Cuspidor | Integrated |
Control Panel | Touchscreen control panel |
Air/Water Syringe | Integrated |
Power Supply | 230 V, 50/60 Hz |
Operating light | LED operating light |
Multimedia integration | Optional |
Hygiene system | Integrated |
Chair Positions | Multiple programmable positions |
Lighting | LED operating light |
Spittoon | Integrated |
Foot Control | Foot control available |
Air Supply | 500 - 800 kPa (5 - 8 bar) |
Water Supply | 200 - 500 kPa (2 - 5 bar) |