Scan eXam™ One
Digital Intraoral Imaging Plate System
User Manual
209643 rev. 10 | 0.805.5080
ENGLISH

Table of Contents

Other manuals for KaVo Scan eXam One

Questions and Answers

  • E
    Elizabeth ChaneyAug 13, 2025
    Why are there ghost images or shadows on my KaVo All in One Printer images?
    • C
      crystalrossAug 13, 2025
      Ghost images and shadows can appear on your KaVo All in One Printer images if the imaging plate was exposed twice without processing between exposures, or if more than one image was exposed to the same plate. To resolve this, avoid exposing the imaging plate twice without processing and ensure only one image is exposed to the plate. Also, ensure the imaging plate is properly erased after processing. Check the LEDs during operation and respond to warnings, as the unit monitors erasing LEDs during normal operation, and the application software shows a warning if LEDs are defective.
  • K
    Kristine RobinsonAug 17, 2025
    Why does my KaVo All in One Printer have decreased contrast, shadows, or ghost images?
    • S
      shannondavisAug 17, 2025
      Decreased contrast, shadows, or ghost images in your KaVo All in One Printer output can occur due to several reasons. Ensure the plate is properly shielded from light between exposure and processing by using a protective cover correctly. Verify the hygiene bag is sealed properly and that you are using genuine hygiene accessories. Proper storage of imaging plates and correct X-ray dosage are crucial; avoid excessively high X-ray doses. The imaging plate should not be exposed to ultraviolet (UV) radiation. Use a dedicated imaging plate storage box to avoid background radiation. Store the plate away from the X-ray unit. Perform initial erasing for the plate(s) if they have been stored in dark and/or near X-ray unit.
  • C
    Charles LopezAug 21, 2025
    What to do if my KaVo All in One Printer images are unsharp or blurred?
    • E
      Evan SpencerAug 21, 2025
      If your KaVo All in One Printer images are unsharp or blurred, and you notice motion artifacts, it could be due to patient or X-ray cone movement during exposure. Check the exposure procedure to ensure stability. Also, ensure the stability of the X-ray unit. If a long exposure time was used, try using a shorter exposure time, increasing kV if necessary to compensate for the effect of the shorter exposure time.
  • F
    Frank DeleonAug 25, 2025
    Why does my KaVo All in One Printer produce white or grey dots, spots, or stains in images?
    • J
      Jenny LawsonAug 26, 2025
      White or grey dots, spots, or stains in images from your KaVo All in One Printer are often due to dust or stains on the imaging plates, or any extra particle on top of the active sensitive surface of the plate. To resolve this, clean the plate(s). If cleaning doesn't help, consider replacing the plate.
  • T
    Tammy CortezAug 28, 2025
    How to correct cone cut on KaVo All in One Printer images?
    • G
      Gary SawyerAug 28, 2025
      A cone cut on KaVo All in One Printer images happens when the X-ray beam exposes only part of the imaging plate surface. To fix this, check the exposure procedure. Also, use a proper holder during exposure.
  • S
    Shawn MooreSep 1, 2025
    What causes geometry distortion in KaVo All in One Printer images?
    • A
      Ashley ThompsonSep 1, 2025
      Geometry distortion in KaVo All in One Printer images is caused by improper patient positioning. Use proper holders to avoid this issue.
  • A
    Amy HillSep 5, 2025
    What causes wearing of the imaging plate on KaVo All in One Printer?
    • M
      Maria McintyreSep 5, 2025
      Wearing of the imaging plate on your KaVo All in One Printer can be caused by scratches, spots, dots (white or gray) or any visible pattern, moisture, or improper cleaning. Clean the plate(s), using ONLY >70% ETHANOL. Replace the plate if cleaning does not help.
  • A
    Angela RossSep 8, 2025
    Why is there a circular shape on my KaVo All in One Printer image?
    • L
      Leah JohnsonSep 8, 2025
      A circular shape on the image from your KaVo All in One Printer indicates that the imaging plate has been exposed from the wrong side. Ensure the imaging plate is exposed from the correct side.

Summary

2 Basic use

2.2 Positioning and exposure

Guidelines for optimal plate positioning and X-ray exposure settings.

2.3 Processing the imaging plates

Procedure for inserting and processing imaging plates through the scanner.

3 Advanced use

3.1 CLINIVIEW Setup Options

Configure Scan eXam™ One settings using the CLINIVIEW software interface.

3.2 DTX Studio Core Settings

Configure Scan eXam™ One settings via the DTX Studio™ Core software.

5 Introduction to imaging plate technique

5.3 Processing

Detailed explanation of the imaging plate scanning and processing mechanism.

6 Installation of the imaging plate system

6.2 Connecting the unit to a network

Instructions for connecting the Scan eXam™ One to a local area network.

6.3 Install the Application software

Steps for installing the necessary software for system operation.

7 Troubleshooting

7.1 Error images

Diagnose image issues like ghost images, shading, or decreased contrast.

7.2 Error messages

Understand unit error codes and their meanings for problem diagnosis.

8 Other information

9 Technical specifications

10 Symbols and labeling

10.3 Warnings and precautions

Critical safety warnings and precautions for using the device.

KaVo Scan eXam One Specifications

General IconGeneral
BrandKaVo
ModelScan eXam One
CategoryDental equipment