9 Technical specifications
KaVo Scan eXam One 63
9.5 Electromagnetic Compatibility
(EMC) tables
NOTICE! Medical electrical equipment needs
special precautions regarding EMC and needs to be
installed according to EMC information.
IEC60601-1-2 ed4 testing has verified that
electromagnetic interference stimulus has no effect
to safety critical functionality of the device. This
includes image readout and erasing.
If abnormal performance is observed, such as
degradation of essential performance in form of
reduction of line pair resolution, additional
measures may be necessary, such as re-orienting
or relocating the device.
The device is suitable for use in both at professional
healthcare (hospitals/large clinics) facility
environment and home healthcare (clinics in
domestic establishments and those directly
connected to the public low-voltage power supply)
environment.
Exceptions for professional healthcare facility
environment: Not to be used or installed near
active HF SURGICAL EQUIPMENT and the RF
shielded room of an ME SYSTEM for magnetic
resonance imaging, where the intensity of EM
DISTURBANCES is high.