9 Technical specifications
64 KaVo Scan eXam One
Table 1: Electromagnetic emissions IEC 60601-1-2 Ed4
The device is suitable for use in the specified electromagnetic environment. The pur-
chaser or user of the device should assure that it is used in an electromagnetic en-
vironment as described below:
Emissions Test Compliance Electromagnetic Environment
Radio-Frequency
Emissions CISPR11
Group 1 The device uses RF energy only for its internal func-
tion. Therefore, the RF emission is very low and not
likely to cause any interference in nearby electronic
equipment.
Radio-Frequency
Emissions CISPR11
Class B The device is suitable for use in both at professional
healthcare (hospitals/large clinics) facility environ-
ment and home healthcare (clinics in domestic es-
tablishments and those directly connected to the
public low-voltage power supply) environment.
Exceptions for professional healthcare facility envi-
ronment:
Not to be used or installed near active HF SURGI-
CAL EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where
the intensity of EM DISTURBANCES is high.
Harmonic emissions
IEC61000-3-2
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies