9 Technical specifications
KaVo Scan eXam One 65
Table 2: Electromagnetic immunity IEC 60601-1-2 Ed4
The device is suitable for use in the specified electromagnetic environment. The
purchaser or user of the device should assure that it is used in an electromagnetic
environment as described below:
Immunity
Test
IEC60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge
(ESD)
IEC61000-4-2
± 8 kV for contact
discharge
±2, 4, 8, 15 kV
for air discharge
± 8 kV for con-
tact discharge
±2, 4, 8, 15 kV
for air discharge
Floors are wood, concrete, or ce-
ramic tile, or floors are covered
with synthetic material and the rel-
ative humidity is at least 30 per-
cent.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines (100
kHz)
±1 kV for input/
output lines
(100kHz)
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality is that of a
typical commercial and/or hospital
environment
Surge
IEC61000-4-5
± 0.5, 1 kV differ-
ential mode
± 0.5, 1, 2 kV
common mode
± 0.5, 1 kV dif-
ferential mode
Mains power quality is that of a
typical commercial and/or hospital
environment.
Voltage dips,
short interrup-
tions and volt-
age variations
on power sup-
ply input lines
IEC61000-4-11
-0% U mains; 0,5
cycle At 0°, 45°,
90°, 135°, 180°,
225°, 270° and
315°
-0% U mains; 1
cycle, At 0°-70%
U mains; 25/30
cycles At 0°
-0% U mains;
250/300 cycle At
0°
-0% U mains;
0,5 cycle At 0°,
45°, 90°, 135°,
180°, 225°, 270°
and 315°
-0% U mains; 1
cycle, At 0°-70%
U mains; 25/30
cycles At 0°
-0% U mains;
250/300 cycle At
0°
Mains power quality is that of a
typical commercial and/or hospital
environment. If the user of the de-
vice requires continued operation
during power mains interruptions,
it is recommended that the device
is powered from an uninterruptible
power supply.
Power frequen-
cy (50/60Hz)
magnetic field
IEC 61000-4-8
30A/m 30 A/m Power frequency magnetic fields
are at levels characteristic of a typ-
ical location in a typical commercial
and/ or hospital environment. The
device shall not be used closer
than 15cm to sources of 50/60Hz
magnetic field.
NOTICE! UT is the a.c. mains voltage prior to application of the test level.