9 Technical specifications
66 KaVo Scan eXam One
Table 3: RF immunity of non-life-support equipment or system IEC
60601-1-2 ed.4
The device is suitable for use in the specified electromagnetic environment. The pur-
chaser or user of the device should assure that it is used in an electromagnetic envi-
ronment as described below:
Immunity
Test
IEC60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Conducted RF
IEC 61000-4-6
3V 150kHz to 80
MHz, 6 V in ISM
and amateur radio
bands between
0,15 MHz and 80
MHz according to
table 60601-1-2.
Ed4. Table 5/8.
3V 150kHz to 80
MHz, 6 V in ISM
and amateur ra-
dio bands be-
tween 0,15 MHz
and 80 MHz ac-
cording to table
60601-1-2. Ed4.
Table 5/8.
Portable and mobile RF communi-
cations equipment are used no
closer to any part of the device in-
cluding cables, than the recom-
mended separation distance
calculated from the equation ap-
propriate for the frequency of the
transmitter.
Recommended Separation
Distance:
150kHz-80MHz
80MHz to 800 MHz
800 MHz to 2,7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation distance
in meters (m). Field strengths from
fixed RF transmitters, as deter-
mined by an electromagnetic site
survey,* are less than the compli-
ance level in each frequency
range.**
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Radiated RF
IEC 61000-4-3
10 V/m 80 MHz to
2,7 GHz Immunity
to proximity fields
from RF wireless
communication
equipment, levels
according to table
60601-1-2 ed4 ta-
ble 9.
10 V/m 80 MHz to
2,7 GHz Immuni-
ty to proximity
fields from RF
wireless commu-
nication equip-
ment, levels
according to table
60601-1-2 ed4
table 9.