3 Advanced use
KaVo Scan eXam One 15
3.1.1.3 Image Processing
Noise filtering makes images smoother when they
are taken at short exposure times.
There are two options available:
Classic mode offers traditional noise filtering
algorithms that has been applied to all previous
models of imaging plate systems.
Progressive mode applies another algorithm that
reduces noise while efficiently retaining image
clarity. The progressive algorithm requires
appropriate exposure level for efficient
performance.
Ensure that the exposure level indicator is OK.
3.1.1.4 Retrieve last image
If the last image processed is not transferred to the
workstation because of a network, communication,
workstation or software failure, the last image
processed can be retrieved.
NOTICE! The LAST processed image can only be
retrieved if the unit is left on. If the unit is switched
off the image is lost.
To retrieve the last processed image:
1. Correct the problem that caused the communi-
cation failure. When the connection between the
unit and the workstation is re-established the
last processed image is automatically trans-
ferred to the workstation.
2. Workstation: If the image is not automatically
transferred to the workstation, select the Setup
> Scanner page from the imaging application
software your are using.