Instructions for use MULTIflex Coupling 465 RN 0.553.1600, MULTIflex LUX Coupling 465 LRN 0.553.1550
3 Product description | 3.2 Technical Specifications
12 / 28
3.2 Technical Specifications
Product 465 RN 465 LRN
Voltage range of the
high-pressure lamp
3.0 to 3.35 V DC (KaVo
recommends 3.2 V DC)
Power of the high-pres-
sure lamp
max. 2.5 Watt
Connector in accordance with DIN EN ISO 9168 type 3
Spray control with regulating ring for Spray control
Anti-retraction valve with Anti-retraction valve
Can be attached to all KaVo (LUX) instruments with original MULTIflex
connector
NOTE
Please comply with the Instructions for Use of the product to be connected to
the MULTIflex coupling. Check the suitability of the product for operation with
the MULTIflex coupling.
For 465 LRN only
NOTE
Operation on KaVo units:
4 1042: on the unit, set the cold light intensity to the lowest intensity level
and increase the brightness level maximally to intensity level 4.
4 1065 / 1060: If the KaVo MULTIflex LUX coupling is used on these units,
have a technician adjust the voltage.
4 E80 / E70 / E50 / 1058 / 1080 / 1066 / 1062: The KaVo MULTIflex LUX
coupling can be used on these units with any further measures being re-
quired.
NOTE
Mixed operation KaVo MULTI LED lamp / high-pressure lamp:
Due to design features, mixed operation is not permitted on the following
KaVo units:
1065 / 1063 / 1060 / 1061 / 1059 / 1057
These devices must be fully converted to the new LED. Adaptation of the cold-
light voltage by a service technician is required.
NOTE
Operation on third-party units:
For operation on KaVo-approved third-party devices, see www.kavo.com,
item LED compatibility.