4.0 Operation
4.1 Powering Up
NOTE: The manufacturer’s recommendation is for the operator to wear nitrile or other
gloves during basic operation of the KaVo NOMAD PRO 2.
Press and release the Power button to turn on KaVo NOMAD PRO 2. An audible
double tone and an active display panel indicate the device has power.
Technique factors are redisplayed according to what was displayed when the device was
turned off, along with the last saved time setting.
4.2 Lock/Unlock the X-ray
The X-ray lock and unlock
functionality can be activated to
prevent unauthorized use when
the KaVo NOMAD PRO 2 is idle or
stored. This functionality may also
be used for training purposes.
To lock the X-ray, first press and
hold the Select button while
pressing the Patient button.
The device shifts into an X-rays
Locked state, and the buttons can
be released. Repeat the process to
unlock the X-ray device.
NOTE: If the Select button is held down for longer than 3 seconds before the Patient
button is pressed, the Menu mode will activate; if the Select button is pressed and
released before the Patient button is pressed, the Save menu is displayed.
4.3 Ensuring the Right Exposure Time is Set
When power is turned off, the most recent setting for the exposure time is stored in the
memory and redisplays when power is turned back on. However, if the battery is replaced,
the display is reset to the default value.
To change exposure settings press each of the buttons to toggle through the choices:
1) the Patient Size (adult or child)
2) the Image Receptor (film, phosphor plate, or sensor)
3) the Tooth Group (anterior, posterior, or bitewing)
X-ray unlocked X-ray locked