Examples from factory default settings:
Child, Sensor, and Anterior = 0.09sec
Adult, Film, and Posterior = 0.38sec
When necessary, use the Increase or Decrease buttons to adjust the time in 0.01 second
increments. Adjusted exposure settings may be saved, replacing factory settings (see
Section 4.9 Technique Factors Settings and Adjustments).
4.4 Ready the Device
To prevent accidental radiation exposure, properly position the patient and operator
before readying the KaVo NOMAD PRO 2.
Press and release the trigger once to ready the device. The device will briefly display an
Enabling X-rays message before indicating Ready.
NOTE: As a safety precaution, if the trigger is held longer than one second or pulled a
second time while still in the Enabling X-rays state, the device will not transition to the
Ready state. (This is to prevent an accidental activation of the device by unintended
triggering.)
The illuminated green LED, the
message Ready on the display panel,
and a double tone alert confirm that
KaVo NOMAD PRO 2 is prepared
to fire X-rays. The Ready state
continues until either an exposure is
initiated or timeout occurs after 15
seconds of inactivity (accompanied by
a double tone and the return to the
at-rest state).
While the device is in the Ready
state, any changes to the settings
(pressing any user interface panel
button) will end the state.
Enabling X-rays Alert
Ready Alert