32
Table 201
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The PREVENA PLUS™ 125 Therapy Unit is intended for use in the electromagnetic environment specified below.
The customer or the end user of the PREVENA PLUS™ 125 Therapy Unit should assure that it is used in such an
environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF emissions - CISPR 11
(Radiated & Conducted)
Group 1 The PREVENA PLUS™ 125 Therapy Unit uses RF energy only
for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions - CISPR 11
(Radiated & Conducted)
Class B The PREVENA PLUS™ 125 Therapy Unit is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage flucuations / flicker
emissions
IEC 61000-3-3
Complies