35
Table 205
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Test
Frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
Power
(W)
Distance
(m)
Immunity Test
Level (V/m)
385
380 - 390 TETRA 400
Pulse modulation
b
18 Hz
1,8 0,3 27
450
430 - 470
GMRS 460,
FRS 460
FM
c
±5 kHz deviation
1 kHz sine
2 0,3 28
710
704 - 787
LTE Band
13, 17
Pulse modulation
b
217 Hz
0,2 0,3 9745
780
810
800 - 960
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
b
18 Hz
2 0,3 28
870
930
1 720
1 700 -
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3
4, 25; UMTS
Pulse modulation
b
217 Hz
2 0,3 28
1 845
1 970
2 450
2 400 -
2 570
Bluetooth
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
b
217 Hz
2 0,3 28
5 240
5 100 -
5 800
WLAN 802.11
a/n
Pulse modulation
b
217 Hz
0,2 0,3 9
5 500
5 785
NOTE: If necessary to achieve the Immunity Test Level, the distance between the transmitting antenna and the ME
Equipment or ME System may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50% duty cycle square wave signal.
c
As an alternative to FM modulation, a 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.