61
Table 24. CALibrate command summary
Command Description
:CALibration
:PROTected
:LOCK
:SWITch?
:SAVE
:DATA?
:DATE “<string>”
:DATE?
:NDUE “<string>”
:NDUE?
:LLEVel
:SWITch?
:STEP <Step #>
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
:STEP?
:CALCulate
Calibration root command.
All commands in this subsystem
are protected by the CAL
switch.
Lock out calibration (opposite
of enabling cal with CAL
switch).
Request comprehensive CAL
switch state. (0 = locked;
1 = unlocked)
Save cal constants to
EEPROM.
Download cal constants from
2001.
Send cal date to 2001.
Request cal date from 2001.
Send next due cal date to
2001.
Request next due cal date
from 2001.
Low-level calibration sub-
system.
Request low-level CAL
switch state. (0 = locked;
1 = unlocked)
20V AC at 1kHz step.
20V AC at 30kHz step.
200V AC at 1kHz step.
200V AC at 30kHz step.
1.5V AC at 1kHz step.
0.2V AC at 1kHz step.
5mV AC at 100kHz step.
0.5mV AC at 1kHz step.
+2V DC step.
-2V DC step.
0V DC step.
20mA AC at 1kHz step.
+0.2A DC step.
+2A DC step.
2V AC at 1Hz step.
Request the current calibra-
tion step.
Calculate low-level cal con-
stants.

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz

Related product manuals