Power budgeting examples
Example 1
This example is for a fully loaded Model 3706-S with Model 3723 cards (all 2 pole mode).
Slot #
Card
Channel relays
closed
Backplane relays
closed
SLOT 1
3723
30
4
SLOT 2
3723
30
4
SLOT 3
3723
30
4
SLOT 4
3723
30
4
SLOT 5
3723
30
4
SLOT 6
3723
30
4
This produces the following power consumption:
Slot #1 Power Consumed =
700
+
30 x 100
+
4 x 100
=
4100
Slot #2 Power
Consumed =
700
+
30 x 100
+
4 x 100
=
4100
Slot #3 Power Consumed =
700
+
30 x 100
+
4 x 100
=
4100
Slot #4 Power Consumed =
700
+
30 x 100
+
4 x 100
=
4100
Slot #5 Power Consumed =
700
+
30 x 100
+
4 x 100
=
4100
Slot #6 Power Consumed =
700
+
30 x 100
+
4 x 100
=
4100
Totals for each bank can then be calculated as follows:
Slot 1
Slot 2
Slot 3
Total
Bank #1 Power Consumed =
4100
+
4100
+
4100
=
12300
Slot 4
Slot 5
Slot 6
Total
Bank #2 Power Consumed =
4100
+
4100
+
4100
=
12300
Result: Since each bank has not exceeded maximum power, this is OK.

Table of Contents

Other manuals for Keithley 3720

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3720 and is the answer not in the manual?

Keithley 3720 Specifications

General IconGeneral
BrandKeithley
Model3720
CategoryMultimeter
LanguageEnglish

Related product manuals