ArcValidator
Operating manual - EN
ArcValidator RC gives the error
message 'It seems that the load
bank has not been calibrated'?
ArcValidator L 650 load bank has
not been calibrated or the cal-
ibration information has dis-
appeared.
The L 650 load bank has to be sent for calibration.
ArcValidator RC gives the error
message 'Invalid DataStick found'.
The USB memory stick connected
to the ArcValidator RC is not an
ArcValidator DataStick or the stick
is corrupted.
Use another DataStick that came with the ArcVal-
idator delivery package or order one from your
Kemppi representative.
ArcValidator RC gives the error
message 'ArcV DataStick cannot
be found', even though the ArcVal-
idator DataStick is connected to
the ArcValidator RC USB con-
nector.
The DataStick may be damaged or
corrupted.
Remove and reconnect the DataStick memory stick
to the USB connector of the ArcValidator RC device.
Remove the ArcValidator RC from the ArcValidator L
650 load bank and wait for the ArcValidator RC to
turn itself off (red indicator light under the back
panel goes out). Connect the ArcValidator RC to the
ArcValidator L 650 load bank.
ArcValidator RC gives the error
message 'Overheat warning'.
ArcValidator L 650 load bank is
overheated and may be damaged,
if you don't let it cool down.
Do not turn off the ArcValidator L 650 load bank, but
let its fans cool the device until the ArcValidator RC
tells that verification can safely be continued.
ArcValidator RC gives the error
message 'No equipment profiles
saved on the DataStick'.
The DataStick memory sticks that
come with the ArcValidator
product do not contain the inform-
ation needed in the verification of
welding equipment, but the
information has first to be copied
to the memory stick using the
ArcValidator computer applic-
ation.
Connect the DataStick to the computer's USB con-
nection. Start the ArcValidator computer application.
Wait for the PC application to find the DataStick and
complete the file transfer (the PC application auto-
matically copies the welding equipment information
to the memory stick). Safely remove the DataStick.
Connect the DataStick to the USB connector of the
ArcValidator RC device.
ArcValidator RC gives the error
message 'Could not read equip-
ment profiles'.
ArcValidator RC cannot read the
welding equipment profile data
from the DataStick.
Connect the DataStick to the computer's USB con-
nection. Start the ArcValidator PC application. Wait
until the computer application finds the DataStick
and completes the data transfer. Safely remove the
DataStick. Connect the DataStick to the USB con-
nector of the ArcValidator RC device.
ArcValidator RC gives the message
'verification notification'.
The measuring accuracy of ArcVal-
idator L 650 load bank has to be
checked once a year, in order to
prevent the verification results
from being faulty because of inac-
curate measurements.
The L 650 load bank has to be sent for calibration.
ArcValidator RC gives the error
message 'It seems that welding
equipment is connected to more
than one wire feed unit'.
The welding system contains
more than one wire feeder, but
automatic verification is only pos-
sible in welding systems with only
one wire feeder.
Disconnect the superfluous wire feeders from the sys-
tem.
ArcValidator RC device gives the
message ’Automatic verification is
possible only when one welding
panel is connected’.
More than one panel has been
connected to the Synergic/M
welding system, but automatic
verification is only possible with
one-panel welding systems.
Remove the superfluous panels from the system.
© Kemppi
33
AVPCRC / 2224