ArcValidator RC gives the message
'verification report could not be
saved'.
ArcValidator RC could not save the
verification results on the
DataStick. The stick may be full,
read-only or corrupted.
Check that the DataStick memory is not full and that
the DataStick is not write-protected.If not, use
another DataStick that came with the ArcValidator
delivery package or order one from your Kemppi rep-
resentative.
ArcValidator RC gives the error
message 'System error in load
bank'.
A serious error has occurred the
ArcValidator L 650 load bank that
prevents the use of the system.
Restart the ArcValidator L 650 load bank.
The start of the measurement dur-
ing verification has issues and/or
the load bank doesn't seem to
provide the required power for
the measurement.
The set ignition simulation mode
in the welding equipment profile
may not be suitable for the weld-
ing equipment in question.
Change the ignition simulation mode for the equip-
ment profile in the ArcValidator computer applic-
ation and transfer the new profile(s) to the
ArcValidator RC using the DataStick.
ArcValidator DataStick and ArcValidator computer application
I connected the DataStick to the
computer, but the computer
application does not find it.
Windows does not regognise the
memory stick or the memory stick
is damaged.
Check whether the memory stick can be seen in the
Windows file management system.Remove the
DataStick and connect it to another USB connector
port.Restart your computer.