9 Adjusting (Calibration)
9.1 Introduction
The accuracy of the balance can be checked and adjusted at any time with the
enclosed adjusting weight.
Take care of stable environmental conditions, see point 12 „Important informations“.
A warming up time of approx. 30 minutes is necessary.
Press and hold
F
and only release it when
CAL
appears.
Press and hold
T.
the immediately touch F for a moment, then also release
T
.
Display
Unit 1 (Unit 2)
Touch
F
for a moment. Display
on 0
flashing.
Don’t touch the weighing plate. Briefly after that appears
on F5
in the display.
Place the adjusting weight carefully on the centre of the weighing plate.
Display flashing
on F.5
then a moment later the weight value of the adjusting
weight will be indicated
.
Remove the adjusting weight, the adjusting is terminated.
In case of a adjusting error or in order to interrupt the adjusting process get out with
S
.
Keep the calibration weight near the balance for daily checking is accuracy.
Important!
Handle carefully the adjusting weight. Protect it against damage and dirt.
Arrange for periodic checking. Keep it near the balance (danger of
confusion).