EW/EG-N/EWB-e-1627
36
6.2.2.1 Add mode
This function is used to increase the counting accuracy by increasing the reference
quantity. It can thereby be avoided that an inadequate reference unit is used, as this
could lead to inaccurate results.
When applying this function, the necessary minimum number of items is automatical-
ly ensured.
Operation
Display
1. Carry out item 1-5 as in
chap. 6.2.2 “Parts counter“.
e.g. place 10 items on the weighing
plate.
2. Operate the key.
The reference weight of the 10 items is
stored.
Counting accuracy can be increased by
carrying out the following steps.
3. Doubling the items to be weighed:
Apply a further 10 items (approx.).
Operate the key
The reference weight of the 20 items is
stored.
Pcs
M on 10
Pcs
M 10
Pcs
M 20