1 Technical data
KERN FCF 3K-4 FCF 30K-3
Readability (d) 0.1 g 1 g
Weighing range (max) 3 kg 30 kg
Taring range (subtractive) 3 kg 30 kg
Reproducibility 0.1 g 1 g
Linearity 0.3 g 3 g
Minimum unit weight at
piece counting
0.2 g 2 g
Warm-up time 2 hours 2 hours
Reference quantities at
piece counting
5, 10, 20,25, 50
Weighing Units Details „Weighing units“, see chap. 8.3
Recom. adjustment weight, not
added (class)
Details about „Selection of
adjustment weight“ see chap. 9.3
3 kg (M1) 30 kg (M1)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C ... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm 270 x 323 x 110
Weighing plate mm 253 x 229
Weight kg (net) 2.6