The animal weighing function can be applied for busy weighings.
During a defined period the mean value of the weighing results is
The more unquiet the weighed item, the longer the period should
be selected.
Keep pressed in the weighing mode until “Unit” is
displayed.
Press repeatedly until „ANL“ will be displayed.
Acknowledge using , the current setting is displayed.
Select desired setting by .
ANL off Animal weighing not active
Acknowledge selection by .
Put weighed load (animal) on weighing pan and press . In
the display runs a „Countdown“.
The average value of the weighing results is displayed and
remains displayed on the screen.
Use to change between animal weighing and normal
weighing.
Press to restart the weighing cycle for animal weighing.