FCF-BA-e-1611 15
Parts counting
During piece counting parts can either be counted into a
container or out of a container.
To count a greater number of
parts the average weight per part has to be determined with a
small quantity (reference quantity).
The larger the reference quantity, the higher the counting
exactness.
High reference must be selected for small parts or
parts with considerably different sizes.
The larger the reference quantity, the more accurate the parts
counting.
The process has four steps:
Tare the weighing container
Determine the reference unit
Original weighing of reference weight
Count the items
In weighing mode press . Reference piece number
5
PCS
appears flashing.
By pressing several times other reference quantities
10, 20, 25 and 50 can be called up. Place as many pieces
to count on the weighing plate as the set reference quantity
requires.
Acknowledge with . The balance is now in parts
counting mode counting all units on the weighing plate.
Return to weighing mode
Press .
Error message „Er 1“
Piece below minimum weight of piece, see chap. 1
Technical data; press and restart Reference
determination.
Taring
The tare vessels can also be used for piece counting. Before
starting piece counting use to tare out the tare
container.