36 TFKB-B-BA-e-2311
10.4.2 Weighing with multiplication factor via the application unit <FFA>
Here you determine with which factor the weighing result (in gram) will be multiplied.
By that way, e.g. a known error factor in the weight determination can be immediately
taken into account.
Select menu setting < unit> and confirm on button.
Use the navigation keys to select the setting < FFA >
and confirm on button.
Enter multiplication factor, numerical input
see chap. 3.2.2, the active digit flashes.
10.4.3 Percent weighing by application unit <%>
The application unit <%> allows to check the weight of a sample in percent, based on
a reference weight.
Select menu setting < unit>.
Place a reference weight which corresponds to 100 %
at the upper display
margin
at the upper display
margin
Acknowledge by button.
Use the navigation keys to select the setting < % > and
confirm on button.
Confirm the flashing weight value of the reference weight
using button.
From now on the weight of the sample will be shown in
percent based on the reference weight

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN FKB and is the answer not in the manual?

KERN FKB Specifications

General IconGeneral
BrandKERN
ModelFKB
Weighing capacity6 kg
Weighing range0 - 6 kg
Readability0.1 g
CalibrationExternal
DisplayLCD
Weighing unitsg, kg, lb, oz
InterfaceRS-232

Related product manuals