10. Error messages
Possible causes / Elimination
Err 1 Wrong date
Enter date in format "yy;mm;dd", see
chap.
7 "F7 da"
Err 2 Wrong time
Enter time in format “hh;mm;ss“, see
chap.
7 "F7 da"
Err 4 Error during zero setting
Zero range exceeded
Check if the balance is not loaded
Err 5 Keyboard error
Improper operation of the balance
Err 6
Value outside the A/D changer
range
Damaged weighing cell
Err 7
Error "Determination of
percentage"
Increase the value to >0.5 d
Err 8 Incorrect adjusting weight
Check value for adjustment weight, see
Err 9
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other
Err 10 No WLAN connection
Check menu setting
Error "Communication log"
Check the communication settings
Err 12
Number of totalizing operations > 99
Capacity of balance exhausted
Err 15
Error "Gravitational constant"
Value out of range 09.xx -1.0xx
Err 17
Taring range exceed or not achieved
Err 19
Unable to initialise zero point
Measuring cell defective / overloaded
Object on weighing pan / contact
Main board defective
- - ol - -
Maximum load exceeded
Reduce load
Check whether the balance has been
--lo--
Underload
Negative weight, check platform and
Fai l h / fai l l
/ fai l
Check value for adjustment weight, see
chap. 1
Repeat adjustment process
Ba lo / lo ba
Capacity of rechargeable battery
exhausted
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.