6.10.6 Position of verification safety seal (only on ARJ/PRJ models appropriate
for verification)
M sign for EU – first calibration
(green M)
Balances which are subject to verification must be taken out of operation
if:
- The weighing result from the balance is outside the operational error limit.
Therefore at regular intervals, load the balance with known test weights
(approx. 1/3 of the maximum load) and compare with the display value.
- The re-verification date has been exceeded.
S Safety seals
(self-sticking seal or lead)
K shield of mark with CE-sign and date
of the year
M sign for EU – first calibration
(green M)