Caring for Your Monitoring System
Store meter in the carrying case provided whenever possible.
Wash and dry hands well after handling to keep the meter and test strips free
of water and other contaminants.
Keto-Mojo
®
GK+ Blood Glucose and β-Ketone Meter is a precision electronic
instrument. Please handle it with care.
Avoid exposing meter and test strips to excessive humidity, heat, cold, dust, or dirt.
Cleaning and Disinfection
Your Keto-Mojo
®
GK+ Blood Glucose and β-Ketone Meter should be cleaned and
disinfected a minimum of once per week. Use only Clorox
TM
Healthcare Bleach
Germicidal Wipes, which has been proven to be safe to use with the Keto-Mojo
®
GK+ Blood Glucose and β-Ketone Meter.
Cleaning is part of your normal care and maintenance and should be performed
prior to disinfection, but cleaning does not kill germs. After use and exposure to
blood, all parts of this kit can potentially transmit infectious diseases. Disinfecting
reduces the risk of transmitting infectious diseases.
Note: If the meter is being operated by a second person who is providing testing
assistance to you, the meter should be cleaned and disinfected prior to use by the
second person.
1. Cleaning Your Meter
Step 1: Take one piece of Clorox™ Healthcare Bleach Germicidal
Wipes (EPA Registration No. 67619-12) from the container.
Step 2: Clean the entire meter surface including front side, back side,
right side and left side.
The meter should be cleaned when it is visibly dirty or at a minimum of once per
week. This cleaning is to prepare the meter surface for a disinfection process.
2. Disinfecting Your Meter
Step 1: After cleaning your meter, take out a new piece of Clorox™ Healthcare
Bleach Germicidal Wipes.