8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 117
Measurement Operations & Settings 4
command: :SENSe[n]:[CHANnel[m]]:POWer:WAVelength?
syntax: :SENSe[n]:[CHANnel[m]]:POWer:WAVelength?[<wsp>MIN|MAX|DEF]
description: Inquires the current sensor wavelength.
parameters: none
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the
minimum programmable value and the maximum programmable value
response: The wavelength as a float value in meters.
example sens1:pow:wav? -> +1.55000000E-006<END>
affects: All power meters
dual sensors: Master and slave channels are independent.
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
description: For all sources, overwrites the current calibration values with the factory-set calibration settings. See
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination on page 118 and
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]? on page 121 for information on
calibrating your return loss module.
parameters: none
response: none
example sens1:ret:cal:fact
affects: All return loss modules
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
description: For the currently selected source, start the calibration and save the calibration values for a defined reflectance reference
measurement. See
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l] on page 121 for
information on setting the return loss value of your reference reflector.
parameters: none