118 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
response: none
example sens1:ret:cal:coll:refl
affects: All return loss modules
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
description: For the currently selected source, start the calibration and save the calibration values for a defined termination reference
measurement. See :SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l] on page 121 for
information on setting the return loss value of your reference reflector.
parameters: none
response: none
example sens1:ret:cal:coll:term
affects: All return loss modules
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
description: Queries the T-value (termination calibration value) for the return loss module
parameters: none
response: Termination calibration value as a float in dB
example sens1:ret:cal:term? -> +6.5000E+001
affects: All return loss modules