8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 119
Measurement Operations & Settings 4
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
description: Returns the the current calibration values
1. monitor diode reference power
2. return loss diode reference power
3. monitor diode parasitics power
4. return loss diode parasitics power.
parameters: Returns power values in W
response: none
example sens1:ret:cal:val
affects: All return loss modules