8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 141
Measurement Operations & Settings 4
command: [:SOURce[n]][:CHANnel[m]]:MODUlation:INTernal[:STATe]?
syntax: [:SOURce[n]][:CHANnel[m]]:MODUlation:INTernal[:STATe]?
description: Returns the internal modulation state.
parameters: none
response: 0
:
1
:
Off
On
example: sour1:modu:int? 0
affects: 81950A
Notes
command: [:SOURce[n]][:CHANnel[m]]:MODUlation:INTernal:SBSControl[:LEVel]
syntax: [:SOURce[n]][:CHANnel[m]]:MODUlation:INTernal:SBSControl[:LEVel] <wsp>
<value>[HZ|KHZ|MHZ|GHZ|THZ]|MIN|MAX|DEF
description: Sets the internal frequency modulation frequency.
parameters: Any value in the specified range (see appropriate User's Guide)
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: None
example: sour1:modu:int:sbsc 0.2GHz
affects: 81950A
Notes