142 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: [:SOURce[n]][:CHANnel[m]]:MODUlation:INTernal:SBSControl[:LEVel]?
syntax: [:SOURce[n]][:CHANnel[m]]:MODUlation:INTernal:SBSControl[:LEVel]?<wsp>
[MIN|MAX|DEF]
description: Returns the internal frequency modulation frequency in Hz.
parameters: Optional:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: The current modulation frequency value or the minimum, maximum, or default modulation frequency value in Hz.
example: sour1:modu:int:sbsc? +2.00000000E+008
affects: 81950A
Notes
command: [:SOURce[n]][:CHANnel[m]]:MODUlation:EXTernal[:STATe]
syntax: [:SOURce[n]][:CHANnel[m]]:MODUlation:EXTernal[:STATe]<wsp>OFF|0|ON|1
description: Sets the state of the external amplitude modulation.
parameters: OFF or 0: off
ON or 1: on
response: None
example: sour1:modu:ext on
affects: 81950A only
Notes

Table of Contents