8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 209
Measurement Operations & Settings 4
Table 9 Generating Output Triggers from Power Measurements
Hard ware
Triggering
Trigger Rearming Software Triggering Data Acquisition Functions
sens:func:stat
trig:outp trig:outp:rearm init:imm init:cont MINMax LOGGing|STABility
DISabled - An output trigger will never be generated.
AVGover ON An output trigger is generated for every new power measurement when the averaging time period
finishes.
Applies for all subsequent data
acquisition functions.
MEASure ON An output trigger is generated for every new power measurement when the averaging time period begins.
Applies for all subsequent data
acquisition functions.
AVGover OFF An output trigger is generated when the averaging time period of
the first power measurement finishes. A further hard ware output
trigger cannot be generated until you send trig:outp:rearm.
An output trigger is generated for
every new power measurement
when the averaging time period
finishes. Applies for all subsequent
data acquisition functions.
MEASure OFF An output trigger is generated when the averaging time period of
the first power measurement begins. A further hardware output
trigger cannot be generated until you send trig:outp:rearm.
An output trigger is generated for
every new power measurement
when the averaging time period
begins. Applies for all subsequent
data acquisition functions.
command: :TRIGger
syntax: :TRIGger<wsp>NODEA|1|NODEB|2
description: Generates a hardware trigger.
parameters: 1 or NODEA:
2 or NODEB:
Is identical to a trigger at the Input Trigger Connector.
Generates trigger at the Output Trigger Connector.
A hardware trigger cannot be effective in the DISabled triggering mode but can be effective in DEFault, PASSthrough or
LOOPback triggering modes, see
:TRIGger:CONF on page 214 for information on triggering modes.