214 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: :TRIGger[n][:CHANnel[m]]:OUTPut:REARm
syntax: :TRIGger[n][:CHANnel[m]]:OUTPut:REARm<wsp>OFF|ON|0|1
description: Sets the arming response of a channel to an outgoing trigger.
See Table 9 on page -209, for information on how this command affects the generation of output triggers using power
measurements.
parameters: A boolean value: OFF or 0: trigger rearming disabled
ON or 1: trigger rearming enabled (default)
If you return to Local control, all modules return to the default setting.
response: none
example: trig1:outp:rearm 1
affects: All power meters, and return loss modules.
dual sensors: Can only be sent to master channel, slave channel is also affected.
command: :TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
syntax: :TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
description: Returns the arming response of a channel to an outgoing trigger.
parameters: none
response: A boolean value: 0: trigger rearming disabled (default)
1: trigger rearming enabled.
example: trig1:outp:rearm? -> 0<END>
affects: All power meters, and return loss modules.
dual sensors: Can only be sent to master channel, slave channel parameters are identical.
command: :TRIGger:CONF
syntax: :TRIGger:CONF<wsp><triggering mode>
description: Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.