94 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: :SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut][:MAGNitude]
syntax: :SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPUT][:MAGNitude]<wsp>
<value>[DB|MDB]
description: Enters a calibration value for a module.
parameters: The calibration factor as a float value
If no unit type is specified, decibels (dB) is implied.
response: none
example: sens1:corr 10DB
affects: All power meters
dual sensors: Master and slave channels are independent.
command: :SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut][:MAGNitude]?
syntax: :SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPUT][:MAGNitude]?
description: Returns the calibration factor for a module.
parameters: none
response: The calibration factor as a float value. Units are in dB, although no units are returned in the response message.
example: sens1:corr? -> +1.00000000E+000<END>
affects: All power meters
dual sensors: Master and slave channels are independent.
command: :SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
syntax: :SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
description: Zeros the electrical offsets for a power meter or return loss module.
parameters: none
response: none