8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 95
Measurement Operations & Settings 4
example: sens1:corr:coll:zero
affects: All power meters and return loss modules
dual sensors: Can only be sent to master channel, slave channel is also zeroed.
command: :SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO?
syntax: :SENSe[n]:[CHANnel[m]]:CORREction:COLLect:ZERO?
description: Returns the status of the most recent zero command.
parameters: none
response: 0:
97
98
99
any other number:
zero succeeded without errors.
zero not done
zero failed
zero absorbed
remote zeroing failed (the number is the error code returned from the operation).
example: sens1:corr:coll:zero? -> 0<END>
affects: All power meters and return loss modules
dual sensors: Master and slave channels are independent.
command: :SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
syntax: SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
description: Zeros the electrical offsets for all installed power meter and return loss modules.
parameters: none
response: none
example: sens:chan:corr:coll:zero:all
affects: All power meters and return loss modules
dual sensors: Command is independent of channel.