8 Service and Maintenance
200 E364xA User’s and Service Guide
Resistive loads (R
L
) For E3640A:
2.7 ,150 W/13.5 ,150 W
For E3641A:
43.8 , 300 W/120 , 300 W
For E3642A:
1.6 , 300 W/8.0 , 300 W
For E3643A:
25 , 300 W/75 , 300 W
For E3644A:
1.0 , 300 W/5.0 , 300 W
For E3645A:
15.9 , 300 W/46.2 , 300 W
-Measure ripple and noise
Current monitoring
resistor (shunt) – R
M1
0.01 , 0.1%
[a]
ISOTEK Co. model A-H
or equivalent
Constant current test setup
Current monitoring
resistor (shunt) – R
M2
0.2 /250 , 0.1%
[a]
ISOTEK Co. model RUG-Z
or equivalent
Measure current rms ripple
and noise
[a] To find the accurate resistance, it is recommended to use a current monitoring resistor after calibration.
Table 8-4 Recommended test equipment

Table of Contents