458 Keysight N9927-90001 User’s Guide
Power Meter (USB) Mode - Option 302
USB Power Meter Settings
You can also use Run / Hold +/- to toggle between Single and Continuous.
Step Detection Mode
The FieldFox supports the Step Detection feature that is present in Keysight
USB power sensors. When enabled, this feature reduces settling time after a
significant step in the measured power. For more information, please refer to
the USB Power Meter documentation.
How to set Step Detection mode
—Press Meas Setup 4.
—Then Step Detection OFF ON
Then choose from the following:
ON Enables step detection (default setting)
OFF Disables step detection
Limits
In Power Meter Mode, Minimum and Maximum limits set a visible range of
acceptable readings between these two values. Readings that are below the
minimum or above the maximum are clearly visible on the power meter display.
How to set Limits
—Press Limit 8.
Then choose from the following:
Min Limit OFF ON
Toggles the display of the Minimum limit ON and
OFF.
Min Limit Value
Sets the Minimum limit power value. Power reading
lower than this will exceed the limit.
Max Limit OFF ON
Toggles the display of the Maximum limit ON and
OFF.
Max Limit Value
Sets the Maximum limit power value. Power reading
higher than this will exceed the limit.

Table of Contents

Other manuals for Keysight Technologies FieldFox N9912A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies FieldFox N9912A and is the answer not in the manual?

Keysight Technologies FieldFox N9912A Specifications

General IconGeneral
BrandKeysight Technologies
ModelFieldFox N9912A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals