Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide 183
Programming the Status Register System
Status Groups
Standard Event Status Enable Register
The Standard Event Status Enable Register lets you choose which bits in the
Standard Event Status Register set the summary bit (bit 5 of the Status Byte
Register) to 1.
4 Execution Error. A 1 in this bit position indicates that an execution error has occurred. Execution errors have
instrument error numbers from −299 to −200.
5 Command Error. A 1 in this bit position indicates that a command error has occurred. Command errors have
instrument error numbers from −199 to −100.
6
User Request Key (Local). A 1 in this bit position indicates that the
Local key has been pressed. This is true even
if the signal generator is in local lockout mode.
7 Power On. A 1 in this bit position indicates that the signal generator has been turned off and then on.
Query: *ESR?
Response: The decimal sum of the bits set to 1
Example: The decimal value 136 is returned. The decimal sum = 128 (bit 7) + 8 (bit 3).
Bit Description
*ESE <data> <data> is the sum of the decimal values of the bits you want to enable.
Example: To enable bit 7 and bit 6 so that whenever either of those bits are set to 1, the Standard Event
Status summary bit of the Status Byte Register is set to 1. Send the command *ESE 192 (128
+ 64).
Query: *ESE?
Response: Decimal value of the sum of the bits previously enabled with the *ESE <data> command.

Table of Contents

Other manuals for Keysight Technologies N5166B CXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5166B CXG and is the answer not in the manual?

Keysight Technologies N5166B CXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5166B CXG
CategoryPortable Generator
LanguageEnglish

Related product manuals