73
Chapter 3 Features and Functions
Amplitude Modulation (AM)
3
Carrier Waveform Shape
• AM carrier shape: Sine or Square waveform. The default is Sine. You
cannot use ramp, pulse, noise, dc, or arbitrary as the carrier
waveform.
• Front-Panel Operation: Press either the or the front-panel
function key.
• Remote Interface Operation:
FUNCtion {SINusoid|SQUare}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Carrier Frequency
The maximum carrier frequency depends on the function selected as
shown below. The default is 1 kHz for all functions.
• Front-Panel Operation: To set the carrier frequency, press the Freq
softkey for the selected function. Then use the knob or numeric
keypad to enter the desired frequency.
• Remote Interface Operation:
FREQuency {<frequency>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Function Minimum Frequency Maximum Frequency
Sine
Square
1 mHz
1 mHz
10 MHz
10 MHz
33210A users guide.book Page 73 Wednesday, July 16, 2008 11:16 AM