15
Chapter 1 Specifications
Keysight 33210A Function / Arbitrary Waveform Generator
4
1
General
Power Supply: CAT II
100 to 240 V
@
50/60 Hz (-5%, +10%)
100 to 120 V
@
400 Hz (± 10%)
Power Consumption: 50 VA maximum
Operating Environment: IEC 61010
Pollution Degree 2
Indoor Location
Operating Temperature: 0 °C to 55 °C
Operating Humidity: 5% to 80% RH,
non-condensing
Operating Altitude: Up to 3000 meters
Storage Temperature: -30 °C to 70 °C
Floating Connector Shields
(Output, Sync, and
Modulation In only): Shields may float at
± 42 V (peak) relative to
earth ground.
[8]
State Storage Memory: Power off state
automatically saved.
Four user-configurable
stored states.
Interface: LAN, LXI-C 10/100
Ethernet, USB 2.0, and
GPIB standard
Language: SCPI - 1993,
IEEE-488.2
Dimensions (W x H x D):
Bench Top: 261.1 mm by 103.8 mm
by 303.2 mm
Rack Mount: 212.8 mm by 88.3 mm
by 272.3 mm
Weight: 3.4 kg (7.5 lbs)
Safety Designed to: UL-1244, CSA 1010,
EN61010
EMC Tested to: MIL-461C, EN55011,
EN50082-1
Vibration and Shock: MIL-T-28800, Type III,
Class 5
Acoustic Noise: 30 dBa
Warm-up Time: 1 hour
LXI Compliance: LXI Class C, V. 1.0
Note: Specifications are subject to change
without
notice. For the latest specifications,
go to the Keysight 33210A product page and
find the 33210A Datasheet.
www.keysight.com/find/33210A
_________________
Footnotes:
1
To upgrade your 33210A to include arbitrary
waveform functionality, go to
www.keysight.com/find/33210U.
2
Add 1/10th of output amplitude and offset
specification per
°C for operation outside
the range of 18
°C to 28 °C.
3
Autorange enabled.
4
DC offset set to 0 V.
5
Spurious output at low amplitude is
-75 dBm (typical).
6
Add 1 ppm / °C (average) for operation out-
side the range of 18
°C to 28 °C.
7
Sine and square waveforms above 3 MHz
are allowed only with an “infinite” burst
count.
8
These shields must all float at the same
potential to avoid instrument damage.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001
du Canada.
N10149