EasyManua.ls Logo

Keysight 34470A - Page 274

Keysight 34470A
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
*STB?
Read Status Byte Query. Queries the condition register for the Status Byte Register group and returns a
decimal value equal to the binary-weighted sum of all bits set in the register.
A condition register continuously monitors the state of the instrument. Condition register bits are updated
in real time; they are neither latched nor buffered.
Parameter Typical Return
(none) +40
Read condition register (with bits 3 and 5 set):
*STB?
l Similar to a Serial Poll, but processed like any other instrument command. This register is read-only;
bits are not cleared when read.
l Returns same result as a Serial Poll, but the "Master Summary" bit (bit 6) is not cleared by *STB?.
l Power cycle and *RST clear all bits in the condition register.
*TRG
Triggers the instrument if TRIGger:SOURce BUS is selected.
Parameter Typical Return
(none) (none)
Return five DC voltage measurements, using a software trigger to start the measurements:
CONF:VOLT:DC
SAMP:COUN 5
TRIG:SOUR BUS
INIT
*TRG
FETCH?
Typical Response:+1.00520000E+01, (5 measurements)
l After setting the trigger source, you must send INITiate to place the DMM in the "wait-for-trigger"
state. The *TRG command is not accepted unless the DMM is in the "wait-for-trigger" state.
Keysight Truevolt Series Operating and Service Guide
275
SCPI Programming Reference

Table of Contents

Related product manuals