[SENSe:]CURRent:{AC|DC}:RANGe {<range>|MIN|MAX|DEF}
[SENSe:]CURRent:{AC|DC}:RANGe? [{MIN|MAX|DEF}]
Selects a fixed measurement range for AC or DC current measurements on the 3 A terminals. For DMMs
with a 10 A range, you cannot select the 10 A range with this command; instead, use[SENSe:]CURRent:
{AC|DC}:TERMinals 10 or CONF:CURRent:{AC|DC} 10.
This parameter setting is not shared between AC and DCmeasurements. The parameters are independent
for AC and DCmeasurements.
On the 34461A, 34465A, or 34470A, use the [SENSe:]CURRent:{AC|DC}:TERMinals command to spe-
cify which current terminals to use for the measurement.
Parameter Typical Return
<range> : {100 µA|1 mA|10 mA|100 mA|1 A|3 A}. Default: AUTO (autorange).
The 34465A/70A have additional DC current ranges of 1 µA and 10 µA.
+1.00000000E-01
Configure AC current measurements using the 1 A range. Make and read two measurements:
CONF:CURR:AC
CURR:AC:RANG 1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+00,+1.04570000E+00
l
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
l If the input signal is greater than can be measured on the specified manual range, the instrument dis-
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
l Unlike CONFigure and MEASure?, this command does not support the 10 A range. Selecting the 10 A
terminal with [SENSe:]CURRent:{AC|DC}:TERMinals forces the specified type of current meas-
urements to use the 10 A range, but it does not affect the values of [SENSe:]CURRent:{AC|DC}:RANGe
or [SENSe:]CURRent:{AC|DC}:RANGe:AUTO.
l Changing the value of this parameter always changes the measurement resolution. The number of
PLCs (NPLC) stays fixed, and this results in a change of resolution in terms of the units of the meas-
urement. For example, if you have a measurement with 1 µA resolution on the 100 mA range, chan-
ging to the 1 A range results in a resolution of 10 µA.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
SCPI Programming Reference
324
Keysight Truevolt Series Operating and Service Guide