2 Base Matrix Configuration
122 Keysight 34934A User’s Guide
8x32 Matrix Configuration
To create an 8x32 matrix, you may connect your field wiring to the 34934A by
three methods:
Terminal Block
You can use terminal block 34934T-002. The terminal block
connects to the 34934A’s two D-Sub connectors; it provides 12 blocks of 12 screw
terminals each, and silk-screened guidance for field wiring the 8x32 matrices.
Follow these sections in order to configure and wire the terminal block:
“34934T-002 Terminal Block: 8x32 Layout” on page 123
“34934T-002 Terminal Block: 8x32 Jumper Configuration” on page 126
“34934T-002 Terminal Block: 8x32 Wiring” on page 129
Configuration Block
You can use configuration block 34934C-001. Follow
these sections in order to configure and wire the configuration block:
“34934C-002 Configuration Block: Layout” on page 132
“34934C-002 Configuration Block: 8x32 Jumper Configuration” on page 134
“34934C-002 Configuration Block: Wiring” on page 139
Direct Wiring
You can wire directly to the module’s D-Sub connectors using
custom cabling. You will have to short specific pins on and between the D-Sub
connectors to create the 8x32 configuration. Review both these sections before
wiring to the module’s D-Subs:
“34934A module D-Sub Connectors: 8x32 Pin Assignments” on page 145
“34934A Module D-Sub Connectors: 8x32 Wiring” on page 147
NOTE
Use of the 34934C-002 precludes use of a piggy-backed 34934T-002 terminal
block for field wiring connections; the combination is not mechanically reliable.

Table of Contents