EasyManua.ls Logo

Keysight 34934A - 8 x64 Matrix Configuration Methods

Keysight 34934A
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Base Matrix Configuration
150 Keysight 34934A User’s Guide
8x64 Matrix Configuration
To create an 8x64 matrix, you may connect your field wiring to the 34934A by
three methods:
Terminal Block
You can use terminal block 34934T-002. The terminal block
connects to the 34934A’s two D-Sub connectors; it provides 12 blocks of 12 screw
terminals each, and an 8x64 removable overlay (supplied with the 34934T-002)
which provides guidance for field wiring the 8x64 matrices. Follow these sections
in order to configure and wire the terminal block:
“34934T-002 Terminal Block: 8x64 Layout” on page 151
“34934T-002 Terminal Block: 8x64 Jumper Configuration” on page 154
“34934T-002 Terminal Block: 8x64 Wiring” on page 157
Configuration Block
You can use configuration block 34934C-001. Follow
these sections in order to configure and wire the configuration block:
“34934C-002 Configuration Block: Layout” on page 160
“34934C-002 Configuration Block: 8x64 Jumper Configuration” on page 162
“34934C-002 Configuration Block: Wiring” on page 168
Direct Wiring
You can wire directly to the module’s D-Sub connectors using
custom cabling. You will have to short specific pins on and between the D-Sub
connectors to create the 8x64 configuration. Review both these sections before
wiring to the module’s D-Subs:
“34934A module D-Sub Connectors: 8x64 Pin Assignments” on page 173
“34934A Module D-Sub Connectors: 8x64 Wiring” on page 175
NOTE
Use of the 34934C-002 precludes use of a piggy-backed 34934T-002 terminal
block for field wiring connections; the combination is not mechanically reliable.

Table of Contents