Expanded-Row Matrix Configuration 3
Keysight 34934A User’s Guide 221
Creating Expanded-Row Matrices Using Terminal Blocks
Four-row matrices use the 34934T-001 terminal block. Eight-row or sixteen-row
matrices use the 34934T-002 terminal block.
Your large matrix will be built as multiples of a base configuration of 4x32, 4x64,
4x128, 8x32, 8x64 or 16x32. Thus, please thoroughly read the appropriate base
configuration section in Chapter 2, “Base Matrix Configuration”, before combining
modules and these terminal blocks to create multiples of your base matrix.
The illustration below shows expansion header locations on the 34934T-001
terminal block; the expansion header layout for the 34934T-002 is the same.
+/- 100V
J24 J25
JUMPERS
ID
CONFIG
JUMPERS
FOR 4X128
AND 4X32
USE OVERLAY
34934T OPT001
1W / 2W / MIXED
1W OPERATION
DUAL (H,L) 4X64
ID
JUMPERS
U
N
U
S
E
D
1 0
2W OPERATION
4X64
C 1
C 2
C 3 C 4 R 1 R 2
H L
H L
H L H L H L H L
C 5
C 7
C 8 C 9 R 3 R 4
C 9 C 10 C 11 C 12 C 13 C 14
H L
H L
H L H L H L H L
C 15 C 16 C 17 C 18 C 19 C 20
C 21 C 22 C 23 C 24 C 25 C 26
H L
H L
H L H L H L H L
C 27 C 28 C 29 C 30 C 31 C 32
R 1 R 2 C 33 C 34 C 35 C 36
H L
H L
H L H L H L H L
R 3
R 4
C 37 C 38 C 39 C 40
C 41 C 42 C 43 C 44 C 45 C 46
H L
H L
H L H L H L H L
C 47 C 48 C 49 C 50 C 51 C 52
C 53 C 54 C 55 C 564 C 57 C 58
H L
H L
H L H L H L H L
C 59 C 60 C 61 C 62 C 63 C 64
4x32, 8x32, 8x64 and 16x32
expand using both J24 and J25
4x64 and 4x128 expand using only J24