EasyManua.ls Logo

Keysight 34934A - Reading Relay Cycle Count

Keysight 34934A
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCPI Programming 4
Keysight 34934A User’s Guide 233
all 34934A modules in the mainframe. The command is executed at power-on or
following a *RST or SYSTem:CPON command.
SYST:MOD:ROW:PROT DEF, ISO
Reading Relay Cycle Count
The DIAGnostic:RELay:CYCles? command queries the cumulative number of
operational cycles (relay cycle count) for any of the 512 crosspoint relays in the
34934A module, since it was last reset. The query can be used for a single channel
or list of channels.
Example:
The following command returns the cycle count at row 8, column 28,
with the module in Slot 4 and configured for 8x64 mode.
DIAG:REL:CYCL? (@4831)

Table of Contents