Base Matrix Configuration 2
Keysight 34934A User’s Guide 175
34934A Module D-Sub Connectors: 8x64 Wiring
When wiring directly to the 34934A’s D-Sub connectors, you must make the
following connections—in addition to row and column connections—to configure
the 8x64 matrix:
Module Configuration:
The 72 connections listed below are required to create
the 8x64 matrix.
R7 1 C5 7 C21 43 C37 8 C53 44
R7 2 C6 11 C22 45 C38 12 C54 46
R8 3 C7 15 C23 49 C39 16 C55 50
R8 4 C8 19 C24 51 C40 20 C56 52
C9 25 C25 73 C41 26 C57 74
ID0 59 C10 27 C26 77 C42 28 C58 78
ID1 39 C11 31 C27 63 C43 32 C59 64
IL1 61 C12 33 C28 65 C44 34 C60 66
IL2 40 C13 35 C29 67 C45 36 C61 68
C14 37 C30 69 C46 38 C62 70
C15 23 C31 71 C47 24 C63 72
C16 29 C32 75 C48 30 C64 76
Description Pin Description Pin Description Pin Description Pin Description Pin Description Pin
To Create This
Column #
Short This Pin
on Bank 1...
...To This Pin on
Bank 2
To Create This
Column #
Short This Pin
on Bank 1...
...To This Pin on
Bank 2
C1 3 5 C33 4 6
C2 7 9 C34 8 10
C3 11 13 C35 12 14
C4 15 17 C36 16 18
C5 1 7 C37 2 8
C6 511C38 612