EasyManua.ls Logo

Keysight 34934A - Page 47

Keysight 34934A
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Base Matrix Configuration 2
Keysight 34934A User’s Guide 47
Placing CONFIG Jumpers for 4x32
The CONFIG jumper area—shown below
with jumpers removed— includes an illustration (on the 4x32 removable overlay)
that shows jumper placement for setting a 4x32 matrix configuration.
There are two 4x2 blocks of CONFIG JUMPER pins on each side of the board’s
center—you must set all four blocks identically.
Place sixteen horizontally positioned jumpers as shown below.
Placing ID Jumpers for 4x32
The ID jumpers allow the 34980A to recognize
the 4x32 matrix configuration upon mainframe bootup. Jumper placement
guidance is provided on the 4x32 removable overlay (below left); the ID jumper
block is shown at right with jumpers removed.
NOTE
In a high temperature-high humidity environment, salt and oil contaminants
transferred from the fingertips to the moveable CONFIG jumpers can affect the
34934A’s published DC Isolation specification if adjacent jumpers are in physical
contact with each other. In applications where DC (channel to channel) isolation >
10 GΩ is required, use clean cotton gloves when handling the CONFIG jumpers.
If cotton gloves are not available, clean the plastic jumper surfaces using a
cotton cloth and 91% isoprolyl alcohol after handling.
CONFIG
JUMPERS
adjacent housings should not touch
adjacent housings
should not touch

Table of Contents