Base Matrix Configuration 2
Keysight 34934A User’s Guide 57
Placing Row Access Jumpers on the 34934C-001 Configuration Block 
The 
jumpers in 3x16 jumper block J8 (at lower board center) determine if row 
connections can be made from the two male D-Sub connectors P1 and P2 (i.e. 
they “enable” the row signal connection points at the D-subs). 
These “row connections” are the traditional access points from field wiring to the 
matrix rows. In practice, your arrangement of test instrumentation and DUT test 
points will determine if you wish to make row connections at the D-subs, at the 
blue 20-pin connectors, or both.
EXTEND ROWS  LEFT (J5) 
EXTEND ROWS RIGHT (J4)
ALSO EXTEND TO (J6), (P2) 
4X32 4X64
4X128
USE
JUMPERS
CONFIG
USE
JUMPERS
CONFIG
4X32 4X64
4X128
J7
J7
J7
ALSO EXTEND TO (J3), (P2) 
If you plan to extend rows 
left, to J5 and J6, place two 
2x8 jumpers as shown at right.
If you plan to extend rows 
right, to J3 and J4, place two 
2x8 jumpers as shown at right.