Base Matrix Configuration 2
Keysight 34934A User’s Guide 73
Placing CONFIG Jumpers for 4x64
The CONFIG jumper area—shown below
with jumpers removed— includes an illustration (on the silk-screen) that shows
jumper placement for setting a 4x64 matrix configuration.
There are two 4x2 blocks of CONFIG JUMPER pins on each side of the board’s
center—you must set all four blocks identically.
1 Place eight horizontally positioned jumpers as shown below.
In a high temperature-high humidity environment, salt and oil contaminants
transferred from the fingertips to the moveable CONFIG jumpers can affect the
34934A’s published DC Isolation specification if adjacent jumpers are in physical
contact with each other. In applications where DC (channel to channel) isolation >
10 GΩ is required, use clean cotton gloves when handling the CONFIG jumpers.
If cotton gloves are not available, clean the plastic jumper surfaces using a
cotton cloth and 91% isoprolyl alcohol after handling.