Service and Repair 1
Keysight DAQ970A/DAQ973A Service Guide 27
To verify or replace the battery
1 Remove AC power cord.
2 Remove the instrument cover (see page 29).
3 Turn the instrument over and remove the battery.
4 Measure the battery voltage using an external DMM.
Hold the red probe on the top (positive (+) side) of the battery and hold the
black probe on the bottom (negative (–) side) of the battery. Replace the
battery if the voltage is below 2.7 V.
If the battery needs to be replaced, ensure that you install the new battery with
its positive (+) side up. Dispose of the old battery in accordance with local
regulations.
5 Replace the instrument cover, apply power, and turn the instrument on; the
instrument will generate an error and beep (this is normal).
6 Set the date and time according to the procedures shown in the DAQ970A/
DAQ973A User’s Guide.
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits
MemoryNon-volatile memory